HumanInsight
Sensors (Basel). 2024 Aug 3;24(15):5027. doi: 10.3390/s24155027.
ABSTRACT
Body temperature must be monitored in patients receiving Hospital-at-Home (HaH) care for COVID-19 and other infectious diseases. Continuous temperature telemonitoring (CTT) detects fever and patient deterioration early, facilitating decision-making. We performed a validation clinical study assessing the safety, comfort, and impact on healthcare practice of Viture®, a CTT system, compared with a standard digital axillary thermometer in 208 patients with COVID-19 and other infectious diseases treated in HaH at the Navarra University Hospital (HUN). Overall, 3258 pairs of measurements showed a clinical bias of -0.02 °C with limits of agreement of -0.96/+0.92 °C, a 95% acceptance rate, and a mean absolute deviation of 0.36 (SD 0.30) °C. Viture® detected 3 times more febrile episodes and revealed fever in 50% more patients compared with spot measurements. Febrile episodes were detected 7.23 h (mean) earlier and modified the diagnostic and/or therapeutic approach in 43.2% of patients. Viture® was validated for use in a clinical setting and was more effective in detecting febrile episodes than conventional methods.
PMID:39124073 | DOI:10.3390/s24155027